Machine-Independent AI for Chest X-Ray Abnormality Classification

Published

Journal of Medical Imaging and Radiation Sciences 54 (3), S12

Authors

Heejun Shin1, Taehee Kim1, Hruthvik Raj1, Muhammad Shahid Jabbar1, Zeleke Desalegn Abebaw1, and Dongmyung Shin1

Affiliations

1AI Engineering Division, Radisen Co. Ltd., Seoul, KOREA, REPUBLIC OF

OBJECTIVE

Many AI methods to detect chest X-ray (CXR) abnormalities have been developed using CXRs from a single machine (e.g., digital radiography (DR) system), reporting reduced diagnostic performance on CXRs from other machines (e.g., computed radiography (CR) systems). Here, we propose a machine-independent AI to address this problem.

MATERIALS & METHODS

8,480 CXRs from a Vietnam hospital were acquired using a DR system (DR1) and annotated by a radiologist as normal or abnormal (e.g., opacity, etc.). 2,696 CXRs from four Indonesian hospitals (IHs) were acquired using different DR or CR systems (DR2 for IH1, CR1 for IH2, CR2 for IH3, CR3 for IH4) and annotated. We trained two AIs (baseline and proposed) using the CXRs from the Vietnam hospital (7,202 for training; 1,278 for testing) to classify CXRs as normal or abnormal. The baseline model was trained by utilizing the conventional CLAHE method. In contrast, the proposed model was trained by perturbating training data based on X-ray hardware-related changes (e.g., sharpness, contrast, and noise).

RESULTS

When we tested both AIs on CXRs from DR1, the diagnostic performance (AUC) was not different (0.962 w/ proposed; 0.962 w/ conventional; p=0.33). For the other CXRs from different machines, the proposed AI outperformed the conventional (DR2: 0.933 w/ proposed, 0.927 w/ conventional, p =0.07; CR1: 0.950 w/ proposed, 0.920 w/ conventional, p=0.004; CR2: 0.936 w/ proposed, 0.910 w/ conventional, p=0.012; CR3: 0.937 w/ proposed, 0.806 w/ conventional, p=0.007).

CONCLUSION

The proposed AI achieved good diagnostic performance (AUC > 0.93) over the CXRs from the different X-ray machines.

Share:

1. Purpose of collecting and using personal information

Sub-One Co. Ltd. collects personal information according to items of personal information collected to receive customer counseling and process inquiries.


2. Items of personal information collected
– Required: Email
– Selections: Company name, mobile phone number


3. Retention and period of use of personal information
Sub-One Co. Ltd. shall be destroyed after storing the information for 3 years after the purpose of personal information collection and use has been achieved.

 

Provided, That if it is necessary to preserve it under the provisions of the relevant statutes, member information shall be kept for a certain period of time prescribed by the relevant statutes.